Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study

Main Article Content

Alamin Y Dhahi
https://orcid.org/0009-0001-6999-3384
Salwan Y Bede
https://orcid.org/0000-0002-0693-933X
Haig Khachadourian
https://orcid.org/0009-0000-0668-6099

Abstract

Background: Measuring implant stability is an important issue in predicting treatment success. Dental implant stability is usually measured through resonance frequency analysis (RFA). Osstell® RFA devices can be used with transducers (Smartpeg™) that correspond to the implants used as well as with transducers designed for application with Penguin® RFA devices (Multipeg™). Aims: This study aims to assess the reliability of a MultiPeg™ transducer with an Osstell® device in measuring dental implant stability. Materials and Methods: Sixteen healthy participants who required dental implant treatment were enrolled in this study. Implant stability was measured by using an Osstell® device with two transducers, namely, Smartpeg™ and Multipeg™. Insertion torque was also measured and recorded as >50 and ≤50 N·cm. Unpaired t-test and Mann–Whitney U test were conducted to assess the relationships of the implant stability values obtained by the two transducers with insertion torque, whereas Pearson and Spearman's correlations were utilized to investigate correlations between the two transducers. Interclass correlation coefficients were applied to assess the reliability between the two transducers. Results: Implant stability measurements (primary and secondary) showed strong positive correlations between Smartpeg™ and Multipeg™. The reliability values between both transducers in primary and secondary implant stability measurements were 0.922 and 0.981, respectively. The use of both transducers revealed higher implant stability measurements for implants inserted with insertion torque > 50 N·cm than those inserted with insertion torque ≤ 50 N·cm. Conclusions: This study demonstrated that the Multipeg™ transducer is reliable in measuring the stability of dental implants using an Osstell® device.   

Downloads

Download data is not yet available.

Article Details

Section

Research Articles

Author Biographies

Alamin Y Dhahi, Department of Oral and Maxillofacial Surgery, College of Dentistry, University of Baghdad, Baghdad, Iraq

Department of Oral and Maxillofacial Surgery, College of Dentistry, University of Baghdad, Baghdad, Iraq

Salwan Y Bede, Department of Oral and Maxillofacial Surgery, College of Dentistry, University of Baghdad, Baghdad, Iraq

Department of Oral and Maxillofacial Surgery, College of Dentistry, University of Baghdad, Baghdad, Iraq

Haig Khachadourian, Specialist Oral Surgeon, Lead Dentist, Primary Health Care Corporation, Doha, Qatar

Specialist Oral Surgeon, Lead Dentist, Primary Health Care Corporation, Doha, Qatar

How to Cite

1.
Dhahi AY, Bede SY, Khachadourian H. Reliability of the multipeg™ transducer in measuring dental implant stability by using a resonance frequency analysis device (Osstell®): An observational clinical study. J Bagh Coll Dent [Internet]. 2023 Sep. 15 [cited 2024 Dec. 25];35(3):58-66. Available from: https://jbcd.uobaghdad.edu.iq/index.php/jbcd/article/view/3455

References

Abdul Lateef T. The effect The Effect Of Treatment Protocol and Implant Dimensions on Primary Stability Utilizing Reso-nance Frequency Analysis. J Bagh Coll Dent. 2017;29:111-16.

Swami V, Vijayaraghavan V, Swami V. Current trends to measure implant stability. J Indian Prosthodont Soc 2016;16:124-30.

Noaman AT, Bede SY. The Effect of Bone Density Measured by Cone Beam Computed Tomography and Implant Dimensions on the Stability of Dental Implants. J Craniofac Surg. 2022;33:E553–7.

Gómez-Polo M, Ortega R, Gómez-Polo C, Martín C, Celemín A, del Río J. Does Length, Diameter, or Bone Quality Affect Primary and Secondary Stability in Self-Tapping Dental Implants?. J Oral Maxillofac Surg. 2016;74:1344–53.

Kittur N, Oak R, Dekate D, Jadhav S, Dhatrak P. Dental implant stability and its measurements to improve osseointegration at the bone-implant interface: A review. Mater Today Proc. 2020;43:1064–70.

Alattar AN, Bede SYH. Does Mixed Conventional/Piezosurgery Implant Site Preparation Affect Implant Stability?. J Craniofac Surg. 2018;29:e472–5.

Ibraheem NS, Al-Adili SS. Assessment of dental Assessment of dental implant stability during healing period and determi-nation of the factors that affect implant stability by means of resonance frequency analysis (Clinical study). J Bagh Coll Dent. 2015;27:109-15.

Becker W, Hujoel P, Becker BE. Resonance frequency analysis: Comparing two clinical instruments. Clin Implant Dent Relat Res. 2018;20:308–12.

Buyukguclu G, Ozkurt-Kayahan Z, Kazazoglu E. Reliability of the osstell implant stability quotient and penguin resonance frequency analysis to evaluate implant stability. Implant Dent. 2018;27:429–33.

Bural C, Dayan C, Geçkili O. Initial stability measurements of implants using a new magnetic resonance frequency analyzer with titanium transducers: An ex vivo study. J Oral Implantol. 2020;46:35–40.

Vandenbroucke JP, von Elm E, Altman DG, Gøtzsche PC, Mulrow CD, Pocock SJ, et al. Strengthening the Reporting of Ob-servational Studies in Epidemiology (STROBE): Explanation and elaboration. Int J Surg. 2014;12:1500–24.

World Medical Association. World Medical Association declaration of Helsinki: Ethical principles for medical research in-volving human subjects. JAMA. 2013;310:2191–4.

Martin Kjaergaard, Vinh GN, Jan B, Joanne P, Paul MW. Comparison of Implant Stability Between Conventional Drilling and Piezosurgical Implant Bed Preparation Techniques. J Oral Implant. 2021;1;49:79–84.

Huwiler MA, Pjetursson BE, Bosshardt DD, Salvi GE, Lang NP. Resonance frequency analysis in relation to jawbone charac-teristics and during early healing of implant installation. Clin Oral Implants Res. 2007;18:275–80.

Andreotti AM, Goiato MC, Nobrega AS, Freitas da Silva EV, Filho HG, Pellizzer EP, et al. Relationship Between Implant Stability Measurements Obtained by Two Different Devices: A Systematic Review. J Periodontol. 2017;88:281–8.

Rowan M, Lee D, Pi-Anfruns J, Shiffler P, Aghaloo T, Moy PK. Mechanical versus biological stability of immediate and de-layed implant placement using resonance frequency analysis. J Oral Maxillofac Surg. 2015;73:253–7.

Chávarri-Prado D, Brizuela-Velasco A, Diéguez-Pereira M, Pérez-Pevida E, Jiménez-Garrudo A, Viteri-Agustín I, et al. In-fluence of cortical bone and implant design in the primary stability of dental implants measured by two different devices of resonance frequency analysis: An in vitro study. J Clin Exp Dent. 2020;12:e242–8.

Herrero-Climent M, Falcão A, López-Jarana P, Díaz-Castro CM, Ríos-Carrasco B, Ríos-Santos JV. In vitro comparative anal-ysis of two resonance frequency measurement devices: Osstell implant stability coefficient and Penguin resonance frequency analysis. Clin Implant Dent Relat Res. 2019;21:1124–31.

Koo TK, Li MY. A Guideline of Selecting and Reporting Intraclass Correlation Coefficients for Reliability Research. J Chiropr Med. 2016;15:155–63.

Zita Gomes R, De Vasconcelos MR, Lopes Guerra IM, De Almeida RAB, De Campos Felino AC. Implant Stability in the Pos-terior Maxilla: A Controlled Clinical Trial. Biomed Res Int. 2017;2017.

Similar Articles

You may also start an advanced similarity search for this article.